ABSTRACT

Background: Minimally invasive sacroiliac joint (SIJ) fusion (SIJF) has become an increasingly accepted surgical option for chronic SI joint dysfunction, a prevalent cause of chronic low back/buttock pain.

Objective: To report clinical and functional outcomes of SIJF using 3D-printed triangular titanium implants (TTI) for patients with chronic SI joint dysfunction.

Methods: A total of 28 subjects with SIJ dysfunction at 8 centers underwent SIJF with 3D TTI and had scheduled follow-up to 6 months (NCT03122899).

Results: Mean preoperative SIJ pain score was 79.1 and mean preoperative Oswestry Disability Index (ODI) was 49.9. At 6 months, pain scores decreased by 51 points and ODI decreased by 23.6 points (both p<0.0001). The proportion of subjects able to perform various back/pelvis-related physical functions with minimal difficulty improved significantly for all activities. Opioid use decreased and physical function, as assessed with three objective tests, improved.

Conclusions: Early results from this prospective multicenter trial confirm that clinical responses to a 3D triangular titanium implant for SIJF are similar to those from prior trials, with improved physical function and decreased opioid use.

Level of Evidence: II

Author Information

  • Patel V - Department of Orthopedics, University of Colorado, Aurora, CO, USA
  • Kovalsky D - Orthopaedic Center of Southern Illinois, Mt. Vernon, IL, USA
  • Meyer SC - Columbia Orthopaedic Medical Group, Columbia, MO, USA
  • Chowdhary A - Neuroscience Institute, Overlake Hospital, Bellevue, WA, USA
  • Lokcstadt H - Bluegrass Orthopedics, Lexington, KY, USA
  • Techy F - ClinTech Center for Spine Health, Johnstown, CO, USA
  • Billys J - FastMD, Odessa, FL, USA
  • Limoni R - BayCare Clinic Orthopedics and Sports Medicine, Green Bay, WI, USA
  • Yuan P - Memorial Orthopaedic Surgical Group, Long Beach, CA, USA
  • Kranenburg A - South Oregon Orthopedics, Medford, OR, USA
  • Cher D - Clinical Affairs, SI-BONE, Inc, Santa Clara, CA, USA
  • Tender G - Department of Neurosurgery, Louisiana State University, New Orleans, LA, USA


Disclosure: The study was funded by SI-BONE (Santa Clara, California), the manufacturer of the iFuse Implant System used in the study. One author of this study (D.C.) is an employee at SI-BONE.

Looking for Answers?

We can help. Tell us what you're looking for below.